Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections.

Gould L, Pan T, Loghin C, et al. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med. 2007;48(7):1112-21.

Save
Print
October 6, 2011
Gould L, Pan T, Loghin C, et al. J Nucl Med. 2007;48(7):1112-21.
View more articles from the same authors.
In this study, 40% of cardiac PET/CT scans showed a false-positive result due to technical errors. The authors discuss methods for avoiding such errors.